Being light and compact, this instrument can be carried anywhere without
effort and is quicker and easier to use than conventional types. A large,
clear meter is provided, reading 0 to 100% LEL (high concentration range)
and 0 to 10% LEL (low concentration range). A single multi-position switch
turns the instrument on, checks battery voltage and selects the sensitivity
range required. An integral extremely low power drain micro-pump provides
a continuous, steady sample flow to give accurate and repetitive measurements.
The standard model is calibrated in % LEL for LP gas (isobutane) and is
provided with additional meter scales of 0 to 1.8% and 0 to 0.18% by volume.
The instrument is direct reading for the gas used for calibration. Calibration
for other specific gas is available on request. Pushbutton meter illumination
is provided for testing at night or in basements, service tunnels and
other poorly lit areas. A variety of probe accessories is available for
use according to atmospheres to be tested. When remote sampling is not
needed, the instrument may be used as a hand-held portable with standard
probe attached direct to the sample intake. A set of 4 UM-3 (size AA)
dry cell batteries gives up to 4 hours of continuous operation or 100
intermittent tests of 3 minutes duration each.
- Compact and Lightweight
- Automatic Sampling
- High Sensitivity
- Accurate and Dependable
- Long Sensor Life
- Fast and Simple Operation
- Easy to maintain
- Ruggedly Constructed.
Special Models
Model XP-311A - this model has an integral alarm circuit. An intermittent
tone sounds and a red LED lamp flashes if gas concentration exceeds the
preset alarm level. When battery operating voltage drops below the required
level, a steady tone sounds to indicate that the batteries require replacement.
Other special Models - Model XP-311 can be supplied with an output for
connection to a recorder unit. Model XP-311 A can be supplied with an
output for connection to either or both of a recorder and a remote alarm
signal unit.
Specifications
| Detection Principle: |
Catalytic combustion on a platinum filament with very low power
drain. |
| Range: |
0-10% and 0-100% LEL combustibles. |
| Calibrating Gas: |
Standard Model - LP Gas (Isobutane). Semi-standard Model: XP-311(M)
- Methane. Calibration for other specific gas is available on request.
|
| Accuracy: |
± 5% of full scale reading. |
| Response Velocity: |
Max. 3 seconds. |
| Alarm Style and Preset Point (Model XP-311A only):
|
Intermittent tone and flashing lamp. Factory preset at 20% LEL,
but available at other point on request. |
| Operation Temperature: |
-200C to +500C (-40F to 1220F). |
| Power Supply: |
4 x UM-3 (size AA) dry cell batteries. |
| Duty Cycle: |
100 intermittent tests of 3 minutes duration each or 4 hours for
continuous operation using standard dry cell batteries |
| Intrinsic Safety: |
Id2 G3. |
| Dimensions: |
190mm H x 84mm W x 40mm D (7.5" x 3.3" x 1.6") |
| Weight: |
700 gr. (1.5 lbs.). |
| Accessories Supplied: |
Standard probe (AT-3A) with filter/moisture trap (DF-4). Sampling
tube 1m long (SH-6-1). Carrying case with shoulder strap. Set of 4
x UM-3 (size AA) dry cell batteries. Filter refills (FE-2; 2 pcs.). |
| Optional Accessories: |
Sampling tube 5m long (SH-6-5) or other length as required (available
up to 20m). Floating Probe (SF-10A) with float and filter/moisture
trap for testing a confined space where there might be water at the
bottom. Cooling-down probe (CF-3A) with cooler and filter/moisture
trap for testing heated atmospheric air. Telescopic probe. Flexible
probe. Remote alarm signal unit (BEZ-1) with 10m long cable. |
|